ORIGINAL ARTICLE
Year : 2020  |  Volume : 11  |  Issue : 4  |  Page : 336-341  

Influence of implant dimensions in the resorbed and bone augmented mandible: A finite element study


1 Department of Dento-alveolar and Maxillofacial Surgery, Dentures Technology, Faculty of Dental Medicine, University of Medicine and Pharmacy “Grigore T. Popa,” Iasi, Romania
2 Department of Implantology, Removable Dentures, Dentures Technology, Faculty of Dental Medicine, University of Medicine and Pharmacy “Grigore T. Popa,” Iasi, Romania
3 Department of Biomedical Sciences, Faculty of Medical Bioengineering, University of Medicine and Pharmacy “Grigore T. Popa”, Iasi, Romania

Date of Submission24-Oct-2019
Date of Decision22-May-2020
Date of Acceptance08-Jul-2020
Date of Web Publication20-Dec-2020

Correspondence Address:
Assistant Prof. SÓnziana Anca Butnaru Moldoveanu
Department of Biomedical Sciences, Faculty of Medical Bioengineering, University of Medicine and Pharmacy “Grigore T. Popa,”-9-16 M. Kogalniceanu Street, 700454 Iasi
Romania
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ccd.ccd_366_19

Rights and Permissions
   Abstract 


Aims: The scope of this study was to analyze the influence of clinically feasible implant diameter and length on the stress transmitted to the peri-implant bone in the case of a resorbed and bone augmented mandible through finite element analysis. Settings and Design: The study was carried out in silico. Subjects and Methods: Resorbed and bone-augmented 3D models were derived from in vivo cone-beam computed tomography scans of the same patient. Corresponding implant systems were modeled with the diameter ranging from 3.3 to 6 mm and length ranging from 5 to 13 mm, and masticatory loads were applied on the abutment surface. Statistical Analysis Used: None. Results: In the bone augmented ridge, maximum stress values in the peri-implant region drastically decreased only when using implants of a diameter of 5 mm and 6 mm. Implants up to 4 mm in diameter led to comparable stress values with the ones obtained in the resorbed ridge, when using the larger implants. The increase of length reduced stress in the resorbed mandible, whereas in the bone augmented model, it led to small variations only in implants up to 4 mm in diameter. Conclusions: It was concluded that bone augmentation provides the optimal framework for clinicians to use larger implants, which, in turn, reduces stress in the peri-implant region. Diameter and length play an equally important role in decreasing stress. Implant dimensions should be carefully considered with ridge geometry.

Keywords: Bone augmentation, bone resorption, finite element analysis, implant dimensions, peri-implant stress


How to cite this article:
Agop Forna D, Forna NC, Butnaru Moldoveanu SA. Influence of implant dimensions in the resorbed and bone augmented mandible: A finite element study. Contemp Clin Dent 2020;11:336-41

How to cite this URL:
Agop Forna D, Forna NC, Butnaru Moldoveanu SA. Influence of implant dimensions in the resorbed and bone augmented mandible: A finite element study. Contemp Clin Dent [serial online] 2020 [cited 2021 Jun 20];11:336-41. Available from: https://www.contempclindent.org/text.asp?2020/11/4/336/304139




   Introduction Top


Dental implants have been proven to be a successful solution in the oral rehabilitation of edentulous patients. Advancements in materials, coating technology, and a wide variety of designs have allowed dental implants to be used in almost all clinical cases.[1] Among the most important factors in choosing the optimal implant are bone quality, density, and dimensions.[2],[3] Implant dimensions are dependent on available bone volume and also have a great influence on the stress values and distribution in the mandibular bone. Certain dimensions of implants may not even be considered in some cases such as bone resorption due to restrictive dimensions of the mandibular bone. When the loss of bone tissue is substantial, bone augmentation can be done, aiming to restore its dimensions, both in height and width, thus allowing the use of a larger implant. Bone augmentation is a surgical procedure that presents certain risks such as postoperative morbidity of the donor site, increased costs of oral rehabilitation, pain, treatment time.[4]

Many current in vivo and in vitro studies have tried to assess the impact of implant dimensions on the stress transmitted to the peri-implant tissue but have yet to reach a clear conclusion on this complex matter.[4],[5],[6],[7],[8] Some studies have reported that a larger implant may lead to smaller stress values in the peri-implant bone.[5],[6],[7] However, other studies show that smaller implants used in resorbed mandibles have a comparable long-term survival rate and stress values to larger implants.[4],[5],[6] Furthermore, there is a lack of studies on the importance of bone geometry and its influence on the implant induced stress at the crestal bone level, since implant dimensions, stress, and bone geometry are all inter-related and should not be studied separately.[4]

Stress values and distribution are important factors for long-term implant survival. High-stress concentration at the peri-implant bone due to excessive implant loading may lead to bone resorption.[8],[9],[10],[11] Bone loss around the implant neck is associated with implant failure. The long-term success of dental implants depends on the ability of the bone tissue to respond positively to implant loading.[11],[12],[13],[14] Physiological loads applied to the implant are transferred to the surrounding tissue. These forces can either improve bone remodeling or produce bone resorption. Several factors are reported to have a significant influence on the homeostasis of the bone tissue: the type and magnitude of loading, implant geometry and materials, the volume and density of bone tissue, and the characteristics of the bone-implant interface.[12],[13],[14],[15],[16],[17],[18],[19] Approximately 50% of implant losses occur due to the loss of bone support.[20] Implant dimensions can influence bone loss due to their impact on stress values generated in the bone tissue.[20]

Thus, the issue is whether or not bone augmentation is necessary to allow for the use of larger implants when smaller ones could potentially perform comparably and avoid preliminary surgical procedures. Finite element analysis (FEA) is an advantageous technique because it allows the exploration of certain parameters through iterative analysis with no ethical implications that would be otherwise difficult or even impossible to achieve under clinical conditions.[21]

To the best of our knowledge, no previous FEA studies have yet to explore the impact of implant dimensions using a real case of the resorbed mandible and the clinically bone augmented version, by assessing implant dimensions in the pre and post bone augmented mandible model.

The scope of this study was to analyze the influence of clinically feasible implant diameter and length on the stress transmitted to the peri-implant bone in the case of a resorbed and bone augmented mandible through FEA.


   Subjects and Methods Top


Two 3D models representing a segment of the human mandible were derived from two separate in vivo cone-beam computed tomography (CBCT) scans belonging to the same patient, performed pre and post bone augmentation. The mandible presented with mandibular atrophy corresponding to a bone field class II (bone height >10 mm, bone crest width 2.5–5 mm) according to the Misch and Judy classification.[22] Vertical and horizontal bone graft augmentation was performed, aiming to restore bone dimensions. Measurements of bone-implant site pre and postaugmentation are presented in [Figure 1].
Figure 1: (a) Image depicts a cone-beam computed tomography scan image made before the bone augmentation and revealed the initial measurements of the mandibular bone. (b) The image depicts a second cone-beam computed tomography scan image made after the bone augmentation which revealed the obtained measurements of mandibular bone

Click here to view


Models were constructed and processed in Autodesk Fusion360 (Autodesk, Inc., San Rafael, CA, USA). Both mandible sections consisted of two macro-structures, a 2 mm thick cortical bone and an internal cancellous bone, as shown in [Figure 2]. Classic tapered threaded implants were modeled. Implant diameter (D) and length (L) were set as input variables. D ranged from 3.3 mm to 6.0 mm, and L ranged from 5.0 mm to 13.0 mm. A conic 10 mm length abutment was used in all simulated cases to allow for adequate comparison, eliminating a possible effect of the abutment length.[23]
Figure 2: Modeled 3D FE of the resorbed mandibular section (a) and bone augmented mandibular section (b)

Click here to view


Simulations were carried out in Simulation Mechanical version 2017 (Autodesk, Inc., San Rafael, CA, USA). A static type analysis with linear, elastic, and homogeneous material properties was chosen for all simulation cases.

For the implant and abutment, the isotropic mechanical properties of Ti-6Al-4V were assigned. The cortical and the cancellous bone were assumed to be isotropic. Material properties assigned for each analysis element are listed in [Table 1].
Table 1: Material properties

Click here to view


Boundary conditions were applied to end surfaces of the mandibular model, fixed in all directions. The contact type between bone and implant was defined to be perfectly bonded. From a clinical perspective, perfectly bonded contact between bone and implant would translate in perfect osseointegration between the two. Studies have shown that introducing a friction coefficient between the implant and surrounding bone as an expression of various degrees of osseointegration may artificially decrease stress in the peri-implant bone.[24] Masticatory type loads were applied on the abutment surface based on the previous work of Himmlová et al.: 114.6 N in the axial direction, 17.1 N in the lingual direction, and 23.4 N in the distomesial direction as illustrated in [Figure 3].[25]
Figure 3: 3D model of the mandibular bone section with loads applied

Click here to view



   Results Top


In the resorbed mandible, values of stress in the cortical and cancellous bone were highest when using the 3.3D/5L implant. The 3.7D/8L implant led to the lowest stress values in the bone tissue, in the resorbed mandible. In the bone augmented ridge, at cortical bone level, the highest value of stress was recorded when using the 3.7D/13L implant and lowest when using the 5D/11.5L. In the cancellous bone, the 5D/11.5L implant led to a decrease in stress as opposed to the 3.7D/11.5L implant which led to an increase in stress. In the bone augmented mandible model, differences in stress values were small between implants of diameter 5 and 6. Stress values obtained in the simulations in both resorbed and bone augmented mandible are shown in [Figure 4].
Figure 4: von Mises stress in resorbed and bone augmented mandible bone when using various implant dimensions

Click here to view


For all simulated cases, stress was concentrated at the crestal bone level, around the neck of the implant. Peak von Mises stress values were situated opposite side of the applied distomesial masticatory loads, as illustrated in [Figure 5]. A more prominent stress concentration at the crestal bone level in the resorbed mandible was obtained when using the 3.7D/5L implant. In the bone augmented mandible, a similar area was observed when using the 4D/13L implant.
Figure 5: Stress distribution at crestal bone level in the resorbed and bone augmented mandible models when using various implant dimensions

Click here to view



   Discussion Top


In this study, analyzing the resorbed and the bone augmented mandible models using various implant dimensions allowed to evaluate bone dimensions relative to stress values. Unlike many prior studies mentioned in the literature, our research has paid specific attention to using a CBCT derived mandible model of a pre and post bone augmentation clinical case, allowing for an adequate analysis of the influence of various implant dimensions on the stress values and distribution in the mandibular bone. Thus, an assessment was made with smaller implants in the resorbed mandible model and with larger implants in the bone augmented model. The overall results indicate that larger implants, both in diameter and in length, placed in a bone augmented mandible are more favourable in terms of stress values. However, while implants with diameter up to 4 mm led to overall similar stress values in the augmented ridge in comparison to some of the smaller implants in the resorbed ridge, stress only decreased considerably when using larger diameter implants of 5 mm and 6 mm.

In the resorbed mandible model, the increase of the implant diameter showed a consistent decrease of stress in both the cortical and cancellous bone. One notable peak of stress, in both of the bone tissues, was obtained when using the 3.7D/5L implant, where even though there was an increase in implant diameter, the insufficient length led to greater stress values. In the bone augmented mandible, results outlined a similar pattern. Although this in accordance with several studies that suggest that an increase in implant diameter decreases stress in surrounding cortical and cancellous bone, our results showed that this is only valid for certain implant dimensions.[6],[30],[31] This may be due to the dependence of the stress values and the distribution on the model geometry, thus the importance of using real clinical models in such FEA.[32] Lee et al. reported in their study that implants of diameter 6 mm showed the most noticeable reduction in stress values.[6]

However, in our study, both 5D and 6D implants could be favourable choices since they resulted in similar stress values. The diameter of implants must be analyzed in conjunction with length because, as our results show, length can help decrease stress in the mandibular bone as in the case of the smaller and larger implants but may lead to peak stresses in some of the diameters used. The length variable showed that its increase led to smaller stress values in the resorbed mandible in both bone tissues while keeping the diameter as a constant. However, in the bone augmented mandible, the length of the implants did lead to some variation of stress values for implants of diameter up to 4 mm and very little variation for implants of diameters 5 and 6 mm. In our study, we hypothetically went to limit of the available bone height, thus including implants of length 13 mm. However, for clinical applications, a well-known safety distance of 2 mm from the inferior alveolar canal needs to be taken into account when choosing the implant length. Some studies have reported that length may be a less influencing parameter than the diameter of the implant.[33],[34] Conversely, Li et al. reported in their study that length was effective in reducing stress in cancellous bone.[8] Also, in regard to length, Guan et al. concluded in their study that a larger implant offers a larger surface area for the load to be applied, thus resulting in less stress distributed to the surrounding bone tissue.[35] Thus, it is important to note that the diameter and length of the implant correlated with bone dimensions need to be optimized to ensure appropriate stress values. Besides geometry characteristics, bone field quality also needs to be taken into consideration.[36],[37] Treatment planning must be thorough and must cater to the specific needs of the patient. For example, without any pro-implant surgical interventions, in implant-supported restorations in fully edentulous patients, the probability of placing sufficient implants is 58.62%.[38] Implants are made of materials that have a higher elastic modulus compared to bone tissue allowing for better load take on. Bone augmentation alone ensures a larger geometry at implant site which allows the bone to be able to better withstand physiological mastication forces but larger implant results in more volume of material with greater mechanical properties to take over the load. The mechanical characteristics of the available bone tissue such as elastic modulus, density, and behaviour under masticatory loads are important, but the size must also be considered.

The distribution of stress was similar in the resorbed and bone augmented ridge. Stress was concentrated at the crestal region, around the neck of the implants. These results are consistent with several studies, showing that this type of stress concentration is not necessarily dependent on the implant dimensions but rather on the mechanical properties of the cortical and cancellous bone and the type of loading applied.[5],[7],[15],[24]

In interpreting the results, we must take into account several limitations. We considered the bone to be isotropic, and the bone-implant interface to be completely osseointegrated. In reality, the bone is anisotropic and there are various degrees of osseointegration at the bone-implant interface.[39] The obtained results of stress values in the bone tissue may be underestimated when lower degrees of osseointegration in certain areas of the implant-bone interface and are associated with higher stress values and changes in stress patterns.[39] In bone augmented mandible case, from a clinical perspective, different types of bone augmentation techniques may lead to better results in the obtained bone geometry and improved osseointegration. For example, the piezosurgery splitting technique allows for simultaneous implant insertion without creating dehiscence and fenestrations.[40] In FEA, these parameters are hard to replicate or to convey in mathematical variables. Implants with different types of the design were not compared. This parameter can greatly influence the connection between the implant and peri-implant bone tissue and the homogeneous or nonhomogeneous mode in which the forces developed at this level are dispersed.[12],[14],[41] Thus, a favorable design can provide the long-term protection of implant-prosthetic restoration. In interpreting the results, we must take into account that while vertical forces are associated with uniform stress distribution along the implant-bone interface, oblique forces cause shear forces and bending moments on the implant, with the stress concentration on the implant neck and bone contact area.[42] The amplitude of masticatory type loading forces in the current literature is characterized by a large variability and is dependent on the patient, type of food consumed, and position of the implant in the oral cavity, values reported ranging from 100 N to 2400 N.[2],[14] Therefore, if the magnitude of the forces that act on the implant is not well defined, optimization of stress values through the selection of the implant dimensions in a resorbed or bone augmented mandible, becomes a major important factor. Furthermore, the mode of application of loads varies, from the axial and horizontal application of loads to loads applied at an angle ranging 15°–30°.[2] Applied loads must describe as best as possible physiological masticatory loads for reliable results. However, in a linear elastic type study with isotropic material properties, increasing the load would only lead to a linear increase of the resulted stress. Hence even though a load value is important to define, in this type of analysis, you would only expect a direct proportionality.[2]


   Conclusions Top


Within the limitations of this study, it was concluded that bone augmentation provides the optimal framework for clinicians to use larger implants, which, in turn, reduces stress in the peri-implant region. Our results showed that in the bone augmented ridge, maximum stress values in the peri-implant region drastically decreased only when using 5D and 6D implants. Implants up to 4D led to stress values, which were similar with some of the ones obtained in the resorbed ridge, when using the larger implants. With the increase in implant diameter, the stress decreased in both cortical and cancellous bone in the resorbed and bone augmented mandible. Length of implants also played an important role in decreasing stress in bone tissue in the case of the resorbed mandible, whereas in the bone augmented model, it led to small variations only in implants up to 4 mm in diameter. Taking into consideration bone geometry, implant length, and diameter, there must be an optimal balance between these parameters, as shown by the results of our study. FEA used in determining the ideal stress formula at the implant-bone interface can be a useful tool when modeling real clinical situations. Further finite element studies using patient-specific data are needed and should be validated with long-term clinical studies.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Diz P, Scully C, Sanz M. Dental implants in the medically compromised patient. J Dent 2013;41:195-206.  Back to cited text no. 1
    
2.
Rittel D, Shemtov-Yona K, Korabi R. Engineering dental implants. Curr Oral Health Rep 2017;4:239-47.  Back to cited text no. 2
    
3.
Baqain ZH, Moqbel WY, Sawair FA. Early dental implant failure: Risk factors. Br J Oral Maxillofac Surg 2012;50:239-43.  Back to cited text no. 3
    
4.
Amine M, Guelzim Y, Benfaida S, Bennani A, Andoh A. Short implants (5-8?mm) vs. long implants in augmented bone and their impact on peri-implant bone in maxilla and/or mandible: Systematic review. J Stomatol Oral Maxillofac Surg 2019;120:133-42.  Back to cited text no. 4
    
5.
Vidya Bhat S, Premkumar P, Kamalakanth Shenoy K. Stress distribution around single short dental implants: A finite element study. J Indian Prosthodont Soc 2014;14:161-7.  Back to cited text no. 5
    
6.
Lee JH, Frias V, Lee KW, Wright RF. Effect of implant size and shape on implant success rates: A literature review. J Prosthet Dent 2005;94:377-81.  Back to cited text no. 6
    
7.
Demenko V, Linetskiy I, Nesvit K, Hubalkova H, Nesvit V, Shevchenko A. Importance of diameter-to-length ratio in selecting dental implants: A methodological finite element study. Comput Methods Biomech Biomed Engin 2014;17:443-9.  Back to cited text no. 7
    
8.
Li T, Hu K, Cheng L, Ding Y, Ding Y, Shao J, et al. Optimum selection of the dental implant diameter and length in the posterior mandible with poor bone quality – A 3D finite element analysis. Appl Math Model 2011;35:446-56.  Back to cited text no. 8
    
9.
Carter DR, Van Der Meulen MC, Beaupré GS. Mechanical factors in bone growth and development. Bone 1996;18:5S-10S.  Back to cited text no. 9
    
10.
Watzek G. Endosseous Implants: Scientific and Clinical Aspects. Chicago, IL: Quintessence; 1996. p. 291-17.  Back to cited text no. 10
    
11.
Brunski JB. Biomechanics of dental implants. In: Block MS, Kent JN, Guerra LR, editors. Implants in Dentistry: Essentials of Endosseous Implants for Maxillofacial Reconstruction. Philadelphia: W.B. Saunders; 1997. p. 63-71.  Back to cited text no. 11
    
12.
Roos-Jansåker AM, Lindahl C, Renvert H, Renvert S. Nine- to fourteen-year follow-up of implant treatment. Part I: Implant loss and associations to various factors. J Clin Periodontol 2006;33:283-9.  Back to cited text no. 12
    
13.
Baggi L, Cappelloni I, Di Girolamo M, Maceri F, Vairo G. The influence of implant diameter and length on stress distribution of osseointegrated implants related to crestal bone geometry: A three-dimensional finite element analysis. J Prosthet Dent 2008;100:422-31.  Back to cited text no. 13
    
14.
Shemtov-Yona K, Rittel D. Fatigue of dental implants: Facts and fallacies. Dent J (Basel) 2016;4:16.  Back to cited text no. 14
    
15.
Dundar S, Topkaya T, Solmaz MY, Yaman F, Atalay Y, Saybak S, et al. Finite element analysis of the stress distributions in peri-implant bone in modified and standard-threaded dental implants. Biotechnol Biotechnol Equip 2016;30:127-33.  Back to cited text no. 15
    
16.
Lekholm U, Zarb G. Patient selection and preparation. In: Brånemark PI, Zarb G, Albrektsson T, editors. Tissue-Integrated Prostheses. Chicago: Quintessence: 1985. p. 199-11.  Back to cited text no. 16
    
17.
Bozkaya D, Muftu S, Muftu A. Evaluation of load transfer characteristics of five different implants in compact bone at different load levels by finite elements analysis. J Prosthet Dent 2004;92:523-30.  Back to cited text no. 17
    
18.
Goiato MC, dos Santos DM, Santiago JF Jr, Moreno A, Pellizzer EP. Longevity of dental implants in type IV bone: A systematic review. Int J Oral Maxillofac Surg 2014;43:1108-16.  Back to cited text no. 18
    
19.
Tonetti MS, Schmid J. Pathogenesis of implant failures. Periodontol 2000 1994;4:127-38.  Back to cited text no. 19
    
20.
Berglundh T, Persson L, Klinge B. A systematic review of the incidence of biological and technical complications in implant dentistry reported in prospective longitudinal studies of at least 5 years. J Clin Periodontol 2002;29 Suppl 3:197-212.  Back to cited text no. 20
    
21.
Huempfner-Hierl H, Schaller A, Hemprich A, Hierl T. Biomechanical investigation of naso-orbitoethmoid trauma by finite element analysis. Br J Oral Maxillofac Surg 2014;52:850-3.  Back to cited text no. 21
    
22.
Misch CE, Judy KW. Classification of partially edentulous arches for implant dentistry. Int J Oral Implantol 1987;4:7-13.  Back to cited text no. 22
    
23.
de Moraes SL, Verri FR, Santiago JF Jr., Almeida DA, de Mello CC, Pellizzer EP. A 3-D finite element study of the influence of crown-implant ratio on stress distribution. Braz Dent J 2013;24:635-41.  Back to cited text no. 23
    
24.
Korabi R, Shemtov-Yona K, Dorogoy A, Rittel D. The failure envelope concept applied to the bone-dental implant system. Sci Rep 2017;7:2051.  Back to cited text no. 24
    
25.
Himmlová L, Dostálová T, Kácovský A, Konvicková S. Influence of implant length and diameter on stress distribution: A finite element analysis. J Prosthet Dent 2004;91:20-5.  Back to cited text no. 25
    
26.
Van Oosterwyck H, Duyck J, Vander Sloten J, Van der Perre G, De Cooman M, Lievens S, et al. The influence of bone mechanical properties and implant fixation upon bone loading around oral implants. Clin Oral Implants Res 1998;9:407-18.  Back to cited text no. 26
    
27.
Menicucci G, Mossolov A, Mozzati M, Lorenzetti M, Preti G. Tooth-implant connection: some biomechanical aspects based on finite element analysis. Clin Oral Implants Res 2002;13:334-41.  Back to cited text no. 27
    
28.
Papavasiliou G, Tripodakis AP, Kamposiora P, Strub JR, Bayne SC. Finite element analysis of ceramic abutment-restoration combinations for osseointegrated implants. Int J Prosthodont 1996;9:254-60.  Back to cited text no. 28
    
29.
Sevimay M, Usumez A, Eskitascioglu G. The influence of various occlusal materials on stresses transferred to implantsupported prostheses and supporting bone: a three-dimensional finite-element study. J Biomed Mater Res B Appl Biomater 2005;73:140-7.  Back to cited text no. 29
    
30.
Ueda N, Takayama Y, Yokoyama A. Minimization of dental implant diameter and length according to bone quality determined by finite element analysis and optimized calculation. J Prosthodont Res 2017;61:324-32.  Back to cited text no. 30
    
31.
Arsalanloo Z, Telchi R, Osgouie KG. Optimum selection of the dental implants according to length and diameter parameters by FE method in the anterior position. Int J Biosci Biochem Bioinforma 2014;4:265-9.  Back to cited text no. 31
    
32.
Kunte VR, Bhoosreddy AR, Bhoosreddy SA, Pandharbale AA, Rajaram Shinde M, Sukhdeo Ahire B. Alveolar bone dimensions of mandibular posterior teeth using cone beam computed tomography: A pilot study. J Contemp Dent 2016;6:9-14.  Back to cited text no. 32
    
33.
Ding X, Liao SH, Zhu XH, Zhang XH, Zhang L. Effect of diameter and length on stress distribution of the alveolar crest around immediate loading implants. Clin Implant Dent Relat Res 2009;11:279-87.  Back to cited text no. 33
    
34.
Iplikçioglu H, Akça K. Comparative evaluation of the effect of diameter, length and number of implants supporting three-unit fixed partial prostheses on stress distribution in the bone. J Dent 2002;30:41-6.  Back to cited text no. 34
    
35.
Guan H, van Staden R, Loo YC, Johnson N, Ivanovski S, Meredith N. Influence of bone and dental implant parameters on stress distribution in the mandible: A finite element study. Int J Oral Maxillofac Implants 2009;24:866-76.  Back to cited text no. 35
    
36.
Meghea DM, Preoteasa CT, Preoteasa E. Attachment systems for implant overdentures. Literature review. Romanian J Oral Rehab 2014;6:51-6.  Back to cited text no. 36
    
37.
Kamel E, Matei MN. Practical issues involved in subtotal edentation rehabilitation. Romanian J Oral Rehab 2016;8:80-9.  Back to cited text no. 37
    
38.
Gumeniuc A. Outcome implant treatment in complete edentulism: A retrospective 5-year follow-up study. Romanian J Oral Rehab 2013;5:99-103.  Back to cited text no. 38
    
39.
Chun HJ, Cheong SY, Han JH, Heo SJ, Chung JP, Rhyu IC, et al. Evaluation of design parameters of osseointegrated dental implants using finite element analysis. J Oral Rehab 2002;29:565-74.  Back to cited text no. 39
    
40.
Nicolae V. Optimizing preimplantary offer trough piezosurgery techniques. Romanian J Oral Rehab 2015;7:84-9.  Back to cited text no. 40
    
41.
Alexandru A, Cimpoesu R, Melian A, Salceanu M. Study on the behavior of dental alloy CoCrWNbMoV in artificial saliva. Rev Chim 2019;70:165-8.  Back to cited text no. 41
    
42.
Zhang G, Yuan H, Chen X, Wang W, Chen J, Liang J, et al. A three-dimensional finite element study on the biomechanical simulation of various structured dental implants and their surrounding bone tissues. Int J Dent 2016;2016:4867402.  Back to cited text no. 42
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5]
 
 
    Tables

  [Table 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
   Subjects and Methods
   Results
   Discussion
   Conclusions
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed518    
    Printed10    
    Emailed0    
    PDF Downloaded53    
    Comments [Add]    

Recommend this journal