ORIGINAL ARTICLE
Year : 2013  |  Volume : 4  |  Issue : 3  |  Page : 338-342  

Comparison of the diagnostic accuracy of direct digital radiography system, filtered images, and subtraction radiography


1 Department of Dentistry, State University of Maringá, Maringá, Paraná, Oral Radiology, Federal University of Sergipe, Aracaju, Sergipe, Brazil
2 Department of Oral Radiology, State University of Maringá, Maringá, Paraná, Oral Radiology, Federal University of Sergipe, Aracaju, Sergipe, Brazil
3 Department of Oral Diagnosis, State University of Maringá, Maringá, Paraná, Oral Radiology, Federal University of Sergipe, Aracaju, Sergipe, Brazil

Date of Web Publication18-Sep-2013

Correspondence Address:
Wilton Mitsunari Takeshita
Department of Dentistry, Federal University of Sergipe, Av. Cláudio Batista. Aracaju, Sergipe
Brazil
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0976-237X.118391

Rights and Permissions
   Abstract 

Background: To compare the diagnostic accuracy of three different imaging systems: Direct digital radiography system (DDR-CMOS), four types of filtered images, and a priori and a posteriori registration of digital subtraction radiography (DSR) in the diagnosis of proximal defects. Materials and Methods: The teeth were arranged in pairs in 10 blocks of vinyl polysiloxane, and proximal defects were performed with drills of 0.25, 0.5, and 1 mm diameter. Kodak RVG 6100 sensor was used to capture the images. A posteriori DSR registrations were done with Regeemy 0.2.43 and subtraction with Image Tool 3.0. Filtered images were obtained with Kodak Dental Imaging 6.1 software. Images (n = 360) were evaluated by three raters, all experts in dental radiology. Results: Sensitivity and specificity of the area under the receiver operator characteristic (ROC) curve (Az) were higher for DSR images with all three drills (Az = 0.896, 0.979, and 1.000 for drills 0.25, 0.5, and 1 mm, respectively). The highest values were found for 1-mm drills and the lowest for 0.25-mm drills, with negative filter having the lowest values of all (Az = 0.631). Conclusion: The best method of diagnosis was by using a DSR. The negative filter obtained the worst results. Larger drills showed the highest sensitivity and specificity values of the area under the ROC curve.

Keywords: Filtered images, digital radiology, digital subtraction radiography, digitized X-rays, interproximal caries


How to cite this article:
Takeshita WM, Iwaki LC, Da Silva MC, Filho LI, Queiroz AD, Geron LB. Comparison of the diagnostic accuracy of direct digital radiography system, filtered images, and subtraction radiography. Contemp Clin Dent 2013;4:338-42

How to cite this URL:
Takeshita WM, Iwaki LC, Da Silva MC, Filho LI, Queiroz AD, Geron LB. Comparison of the diagnostic accuracy of direct digital radiography system, filtered images, and subtraction radiography. Contemp Clin Dent [serial online] 2013 [cited 2019 Mar 24];4:338-42. Available from: http://www.contempclindent.org/text.asp?2013/4/3/338/118391


   Introduction Top


Caries is a dynamic process that develops due to biochemical and ultrastructural changes. It culminates in to typical signs and symptoms, affects approximately 95% of the population, and may lead to tooth loss. Early diagnosis of proximal caries is of great importance so that preventive actions can be taken instead of curative ones. [1],[2]

Early diagnosis of proximal caries is noticeably assisted by radiographic examination. However, detection of tooth decay through radiographs, even with bitewing X-rays, can be difficult, given that, in its early stages, caries shows little loss of mineralized tissue. [3] Digital radiology emerged due to the need for improvement in diagnostic imaging. [4]

The digital system offers features that provide greater dynamism to images, facilitates interpretation and diagnosis of proximal changes. Subtraction radiography is one of these features, which allows visualisation of changes between two images; these changes are usually impossible to se by naked eye. [3],[5] However, the success of this method depends on the standardization of the radiographic images taken at different moments-the a priori and a posteriori registrations. [6],[7],[8],[9],[10],[11],[12],[13],[14]

In addition, the accuracy of diagnosis may also be enhanced by programs that filter the images. These programs can adjust the brightness and contrast, determine the gray level, inverte the shades of gray, and apply pseudocolors. [15],[16],[17] Few studies compared different types of digital images in the diagnosis of changes in the tooth crown. [5],[7],[17]

The present study aimed to compare the diagnostic accuracy of a direct digital radiography (DDR) with complementary metal-oxide-semiconductor sensor (CMOS) system with four types of filtered images and with a priori and a posteriori registrations of digital subtraction radiography (DSR) for the diagnosis of artificially produced defects in a dental crown.


   Materials and Methods Top


The study was approved by the Ethics Committee on Human Research (Protocol 0469.0.093.000-10, Opinion 763/2010) at the State University of Maringα.

The study sample consisted of 20 healthy premolars extracted due to severe periodontal disease or orthodontic indication. The teeth were arranged in pairs in ten blocks of vinyl polysiloxane (Aquasil, Dentsply, USA), simulating the dental arch in the premolar region, with their proximal surfaces in contact. [16] This material was used to keep the teeth in place and, for being pliable, to permit their removal from the blocks to perform the drilling defects on the surfaces.

Each block was radiographed with dental X-ray equipment (70X, Dabi Atlante, Sγo Paulo, Brazil), with exposure of 70 kVp, 8 mA, cylindrical locator, 40-cm focal length, and 0.4-s exposure time. A pilot test, in accordance with As Low As Reasonably Achievable (ALARA) principle, determined the lowest exposure time that provided an image with ideal brightness and contrast. Images were taken with a CMOS sensor (Kodak RVG 6100, Eastman Kodak, Rochester, NY, USA), positioner to the interproximal technique developed to adapt the digital sensor and a 2-cm thick block of wax. The block was placed between the radiation source and the tooth to simulate the soft tissues. Positioning standardization was done by tooth imprinting, with the impression material adhered to the positioner. [18]

Spherical carbide drills (KG Sorensen) were used to make the defects on the proximal surfaces, first with the 0.25-mm diameter bur, followed by 0.5-mm and 1-mm burs, all performed by the same operator. [19],[20],[21] The procedure consisted of removing the teeth from the block, positioning the drill perpendicularly to the surface, and producing the defects in high speed rotation and constant cooling. The teeth were then replaced in the block and radiographed once more.

All images were saved in Tagged Image File Format format. The a posteriori registration used the program Regeemy Image Registration and Mosaicking Version 0.2.43 [22] (DPI-INPE Sγo Jose dos Campos, SP, Brazil and Vision Lab, Electrical and Computer Engineering Department, University of California, Santa Barbara, USA), a free-download program developed by the National Institute for Space Research for remote sensing. Subtraction was performed with Image Tool 3.0 (University of Texas Health Sciences Center, San Antonio, Texas, USA), another free-download program. [23] Filtered images were generated with the algorithms of Kodak Dental Imaging 6.1 (Carestream Health, Rochester, NY, USA).

The resulting images for DDR-CMOS, filtered images (Perio, Negative, Colors 1, and Colors 2), and DSR with a priori and a posteriori registrations are given in [Figure 1].
Figure 1: Types of radiograph images used for diagnostic accuracy analysis

Click here to view


A total of 360 images were generated, which were randomized and evaluated by three dental radiologists. The specialists determined which radiographic method was more precise to detect the drilled defects. Inter- and intra-examiner reliability was verified by weighted Kappa test (K W ). Inter-examiner coefficient was 0.699, and, for intra-examiner, the scores were 0.907 for examiner A, 0.951 for examiner B, and 0.869 for examiner C.

The images were presented on an Intel Core I3 computer (4-GHz RAM, 500-GB hard disk) with a 20-inch LCD monitor (model AOC), with brightness and contrast calibrated and standardized. The evaluators saw the images on a slide show presentation (Microsoft PowerPoint 2003, Microsoft Corp., Instanbul, Turkey), in a dimmed room and with set no time limit. They were asked to use a five-point scale to detect enamel defects, with numbers ranging as follows: (1) Definitely present, (2) probably present, (3) uncertain, (4) probably absent, and (5) definitely absent. [14] The examiners evaluated the images twice with an interval of 1 month between them.

The data were subject to statistical analysis with the use of Receiver Operator Characteristic (ROC) curves.


   Results Top


The 360 images were rated as follows in [Table 1], while the ROC curves for the different imaging methods are shown in [Figure 2]. DSR showed the highest values of sensitivity and specificity of the area under the ROC curve (Az): 0.896, 0.979, and 1.000 with drills of diamaeter 0.25, 0.5, and 1 mm, respectively. In contrast, the lowest values were found for perio filter (0.685, 0.721, and 0.829 with 0.25, 0.5, and 1 mm drills), and for negative filter (0.631, 0.767 and 0.833 with drills 0.25, 0.5, and 1 mm drills). The 1-mm drill showed the highest values, while the 0.25-mm drill showed the lowest values.
Figure 2: ROC curves for different imaging methods, with 0.25, 0.5, and 1 mm drills

Click here to view
Table 1: Radiographic method, drill size, area under the ROC curve (Az) and standard error

Click here to view


Pairwise comparisons between the areas under the ROC curves of the six imaging methods [Table 2] revealed that DSR differed significantly from almost all other methods: With 0.25-mm drill, it differed from the other methods except for DDR-CMOS and Colors 1. Significant differences were found between DSR and the other five methods with 0.5-mm drill, whereas, with 1-mm drill, no differences were found between DSR and perio filter.
Table 2: Comparison between the areas under the ROC curves of different imaging methods

Click here to view



   Discussion Top


Diagnosis of proximal caries is crucial in preventing dental loss, [1],[2] radiographic examination being the best way to evaluate the problem, particularly with the use of bitewing technique. [24] This study used a digital radiography system to obtain X-ray images, given that its features provide greater dynamism to the images, which facilitates diagnosis and interpretation of proximal changes. [4],[25]

Filters developed to help identification of changes [15],[16],[17] are one of the features that enhance the diagnosis of proximal caries. However, few studies have investigated whether filters are, in fact, a resource for diagnosis. This study compared perio filter with three other imaging tools-negative, colors 1, and colors 2, algorithms of Kodak Dental Imaging Software 6.01. In addition, the results of these four imaging methods were compared to those of DDR-CMOS and of a priori and a posteriori DSR registrations. [2],[4],[6],[8]

The comparison between images with filters and DDR-CMOS revealed that colors 1 showed the best diagnostic accuracy. Studies on the use of filters in the diagnosis of caries showed conflicting results. On one hand, Koob et al., [26] and Kositbowornchai et al., [27] found no evidence that digital radiography is more accurate for detecting proximal caries. On the other hand, Akarslan et al., [28] and Sanden et al., [29] found that filtered images had better results than digital images with no filters.

In the present study, DSR was performed with a priori[10],[11],[25] and a posteriori registrations. A priori registrations were taken using a method that could be used in vivo, that is, with a positioner to the bitewing technique and bite impression on vinyl polysiloxane attached to the positioner. For a posteriori registrations, the free-download program Regeemy 0.2.43 [22] was used, a user-friendly program for dentistry researchers and clinicians. [14] Image Tool 3.0 [23] was used for subtraction, which provided six sets of images [Figure 1].

According to Landis and Kock, [30] Kappa test ranges from <0 to 1.00, with almost perfect agreement varing between 1.00 and 0.81 and considerable agreement between 0.80 and 0.61. The results of intra-examiner agreement (examiner A:K W = 0.907; examiner B:K W = 0.951; examiner C:K W = 0.869) revealed that the use of three experts in radiology influenced positively in the evaluation.

ROC curve analysis showed that DSR reached the best results, with all three drills. The comparison between DDR-CMOS and DSR showed similar results to those of Ono et al., [14] who reported more accuracy with DSR. The values of the area under the ROC curve for the DSR with 0.25, 0.5, and 1 mm drills (Az = 0.896, 0.979, and 1.000, respectively) were similar to those obtained by Ferreira et al., [6] and Ricketts et al. [31]

For the DDR-CMOS images, the results were Az = 0.733, 0.780, and 0.850 for 0.25, 0.5, and 1 mm drills, respectively, which gave higher values than those reported by Botenberg et al., [25] (Az = 0.58). However, these authors used decayed tooth, that is, teeth with no pattern of defect. In contrast, the present study's results are similar to those of έlgüy et al ., [21] who also used teeth with artificially produced defects.

Conserning filtered images, colors 1 showed the best performance, whereas negative filters showed the worst results. DDR [32],[33],[34] and DSR typically showed better results and higher AZ values than images with filters.

[Figure 2] shows that all imaging methods had values above the line that separates the graph, which indicates that all methods are appropriate. However, only DRS had values close to 1 [Table 2], values significantly different from the other methods. In addition, the difference between images with filters and DDR was not statistical, which is another indication that DSR is indeed the most accurate method for detecting proximal defects.

Therefore, the results showed that the smallest constribution to the diagnosis of proximal defects was provided by filtered images, whereas DSR contributed the most. However, whether DSR is the most accurate imaging method can be clarified by further in vivo research, particularly, by studies using the two free-download programs, one for DSR (Image Tool 3.0) and the other for a posteriori registrations (Regeemy); these programs can facilitate and broaden the use of DSR.


   Conclusion Top


Within the limitations of the present study, the best method of diagnosis was a DSR. The negative filter obtained the worst results. Larger drills showed the highest sensitivity and specificity values of the area under the ROC curve.

 
   References Top

1.Wenzel A. Digital imaging for dental caries. Dent Clin North Am 2000;44:319-38.  Back to cited text no. 1
[PUBMED]    
2.Wenzel A, Hintze H, Kold LM, Kold S. Accuracy of computer-automated caries detection in digital radiographs compared with human observers. Eur J Oral Sci 2002;110:199-203.  Back to cited text no. 2
[PUBMED]    
3.Janhom A, van Ginkel FC, van Amerongen JP, van der Stelt PF. Scanning resolution and the detection of approximal caries. Dentomaxillofac Radiol 2001;30:166-71.  Back to cited text no. 3
[PUBMED]    
4.van der Stelt P. Better Imaging: The advantages of digital radiography. J Am Dent Assoc 2008;139:7-13S.  Back to cited text no. 4
    
5.Eberhard J, Hartman B, Lenhard M, Mayer T, Kocher T, Eickholz P. Digital subtraction radiography for monitoring dental demineralization. An in vitro study. Caries Res 2000;34:219-24.  Back to cited text no. 5
[PUBMED]    
6.Haiter-Neto F, Ferreira RI, Tabchoury CP, Bóscolo FN. Linear and logarithmic subtraction for detecting enamel subsurface demineralization. Dentomaxillofac Radiol 2005;34:133-9.  Back to cited text no. 6
    
7.Bittar-Cortez JA, Passeri LA, Bóscolo FN, Haiter Neto F. Comparison of hard tissue density changes around implants assessed in digitized conventional radiographs and subtraction images. Clin Oral Implants Res 2006;17:560-4.  Back to cited text no. 7
    
8.Haiter-Neto F, Wenzel A. Noise in subtraction images made from pairs of bitewing radiographs: A comparison between two subtraction programs. Dentomaxillofac Radiol 2005;34:357-61.  Back to cited text no. 8
[PUBMED]    
9.Bottenberg P, Truyen B, Boca C. Digital subtraction technique in dentistry. Rev Belge Med Dent 2007;62:25-32.  Back to cited text no. 9
    
10.Hwang YJ, Fien MJ, Lee SS, Kim TI, Seol YJ, Lee YM, et al. Effect of scaling and root planning on alveolar bone as measured by subtraction radiography. J Periodontol 2008;79:1663-9.  Back to cited text no. 10
[PUBMED]    
11.Goren AD, Dunn SM, Wolff M, van der Stelt PF, Colosi DC, Golub LM. Pilot study: Digital subtraction radiography as a tool to assess alveolar bone changes in periodontitis patients under treatment with subantimicrobial doses of doxycycline. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008;106:e40-5.  Back to cited text no. 11
[PUBMED]    
12.Güneri P, Gögüs S, Tuğsel Z, Ozturk A, Gungor C, Boyacioğlu H. Clinical efficacy of a new software developed for dental digital subtraction radiography. Dentomaxillofac Radiol 2006;35:417-21.  Back to cited text no. 12
    
13.Güneri P, Göğüş S, Tuğsel Z, Boyacioğlu H. Efficacy of a new software in eliminating the angulation errors in digital subtraction radiography. Dentomaxillofac Radiol 2007;36:484-9.  Back to cited text no. 13
    
14.Ono E, Medici Filho E, Faig Leite H, Tanaka JL, De Moraes ME, De Melo Castilho JC. Evaluation of simulated external root resorptions with digital radiography and digital subtraction radiography. Am J Orthod Dentofacial Orthop 2011;139:324-33.  Back to cited text no. 14
[PUBMED]    
15.Alves WE, Ono E, Tanaka JL, Medici Filho TE, de Moraes LC, de Moraes ME, et al. Influence of image filters on the reproducibility of measurements of alveolar bone loss. J Appl Oral Sci 2006;14:415-20.  Back to cited text no. 15
[PUBMED]    
16.Haiter-Neto F, dos Anjos Pontual A, Frydenberg M, Wenzel A. Detection of non-cavitated approximal caries lesions in digital images from seven solid-state receptors with particular focus on task-specific enhancement filters. An ex vivo study in human teeth. Clin Oral Investig 2008;12:217-23.  Back to cited text no. 16
[PUBMED]    
17.Zangooei Booshehry M, Davari A, Ezoddini Ardakani F, Rashidi Nejad MR. Efficacy of application of pseudocolor filters in the detection of interproximal caries. J Dent Res Dent Clin Dent Prospects 2010;4:79-82.  Back to cited text no. 17
[PUBMED]    
18.Behere RR, Lele SM. Reliability of Logicon caries detector in the detection and depth assessment of dental caries: An in vitro study. Indian J Dent Res 2011;22:362.  Back to cited text no. 18
[PUBMED]  Medknow Journal  
19.Chimello DT, Chinelatti MA, Ramos RP, Palma Dibb RG. In vitro evaluation of microleakage of a flowable composite in Class V restorations. Braz Dent J 2002;13:184-7.  Back to cited text no. 19
[PUBMED]    
20.Arias VG, Campos IT, Pimenta LA. Microleakage study of three adhesive systems. Braz Dent J 2004;15:194-8.  Back to cited text no. 20
[PUBMED]    
21.İlgüy M, Dinçer S, İlgüy D, Bayirli G. Detection of artificial occlusal caries in a phosphor imaging plate system with two types of LCD monitors versus three different films. J Digit Imaging 2009;22:242-9.  Back to cited text no. 21
    
22.Fedorov DV, Fonseca LM, Kenney C, Manjunath BS. Reegemy: Automatic image registration and mosaicking system. Available from: http://regima.dpi.inpe.br/[Last accessed on 2013 July].  Back to cited text no. 22
    
23.Dove SB. UTHSCSA ImageTool: Version 3.0 Final is Here. Available from: http://compdent.uthscsa.edu/dig/itdesc.html [Last accessed on 2013 July].  Back to cited text no. 23
    
24.White SC, Pharoah MJ. Oral radiology: Principles and interpretation, 6 th ed. St. Louis: Mosby/Elsevier; 2009. p. 109-51.  Back to cited text no. 24
    
25.Bottenberg P, Jacquet W, Stachniss V, Wellnitz J, Schulte AG. Detection of cavitated or non-cavitated approximal enamel caries lesions using CMOS and CCD digital X-ray sensors and conventional D and F-speed films at different exposure conditions. Am J Dent 2011;24:74-8.  Back to cited text no. 25
[PUBMED]    
26.Koob A, Sanden E, Hassfeld S, Staehle HJ, Eickholz P. Effect of digital filtering on the measurement of the depth of proximal caries under different exposure conditions. Am J Dent 2004;17:388-93.  Back to cited text no. 26
[PUBMED]    
27.Kositbowornchai S, Basiw M, Promwang Y, Moragorn H, Sooksuntisakoonchai N. Accuracy of diagnosing occlusal caries using enhanced digital images. Dentomaxillofac Radiol 2004;33:236-40.  Back to cited text no. 27
[PUBMED]    
28.Akarslan ZZ, Akdevelioğlu M, Güngör K, Erten H. A comparison of the diagnostic accuracy of bitewing, periapical, unfiltered and filtered digital panoramic images for approximal caries detection in posterior teeth. Dentomaxillofac Radiol 2008;37:458-63.  Back to cited text no. 28
    
29.Sanden E, Koob A, Hassfeld S, Staehle HJ, Eickholz P. Reliability of digital radiography of interproximal dental caries. Am J Dent 2003;16:170-6.  Back to cited text no. 29
[PUBMED]    
30.Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics 1977;33:159-74.  Back to cited text no. 30
[PUBMED]    
31.Ricketts DN, Ekstrand KR, Martignon S, Ellwood R, Alatsaris M, Nugent Z. Accuracy and reproducibility of conventional radiographic assessment and subtraction radiography in detecting demineralization in occlusal surfaces. Caries Res 2007;41:121-8.  Back to cited text no. 31
[PUBMED]    
32.Erten H, Uçtasli MB, Akarslan ZZ, Uzun O, Baspinar E. The assessment of unaided visual examination, intraoral camera and operating microscope for the detection of occlusal caries lesions. Oper Dent 2005;30:190-4.  Back to cited text no. 32
    
33.Kositbowornchai S, Basiw M, Promwang Y, Moragorn H, Sooksuntisakoonchai N. Accuracy of diagnosing occlusal caries using enhanced digital images. Dentomaxillofac Radiol 2004;33:236-40.  Back to cited text no. 33
[PUBMED]    
34.Li G, Yoshura K, Welander U, Shi XQ, McDavid WD. Detection of approximal caries in digital radiographs before and after correction for attenuation and visual response. An in vitro study. Dentomaxillofac Radiol 2002;31:113-6.  Back to cited text no. 34
    


    Figures

  [Figure 1], [Figure 2]
 
 
    Tables

  [Table 1], [Table 2]


This article has been cited by
1 Diagnostic Accuracy of Image Enhancement in Intra-Oral Direct Digital Radiography in the Assessment of Interproximal Caries
Farzad Esmaeili,Teymour Abbasi,Nazli Rabienejad,Shabnam Seyedzadeh Sabounchi
Avicenna Journal of Dental Research. 2016; Inpress(Inpress)
[Pubmed] | [DOI]
2 Accuracy of Digital Subtraction Radiography in the Detection of Vertical Root Fractures
Polyane Mazucatto Queiroz,Helena Aguiar Ribeiro Nascimento,Thais Diniz Jacome da Paz,Felipe Nogueira Anacleto,Deborah Queiroz Freitas
Journal of Endodontics. 2016;
[Pubmed] | [DOI]
3 In VitroEvaluation of Proximal Carious Lesions Using Digital Radiographic Systems
Mayana Soares Vieira,Caroline Parente Ribeiro Nogueira,Marcos André dos Santos Silva,José Roberto de Oliveira Bauer,Etevaldo Matos Maia Filho
The Scientific World Journal. 2015; 2015: 1
[Pubmed] | [DOI]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Add to My List *
* Registration required (free)  

 