ORIGINAL ARTICLE
Year : 2013  |  Volume : 4  |  Issue : 3  |  Page : 313-318  

The effect of pouring time on the dimensional accuracy of casts made from different irreversible hydrocolloid impression materials


Department of Prosthodontics, Government Dental College and Hospital, Patiala, Punjab, India

Date of Web Publication18-Sep-2013

Correspondence Address:
Supneet Singh Wadhwa
Department of Prosthodontics, Christian Dental College, C.M.C. and Hospital, Ludhiana, Punjab
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0976-237X.118368

Rights and Permissions
   Abstract 

Aims and Objectives: To determine the time dependent accuracy of casts made from three different irreversible hydrocolloids. Materials and Methods: The effect of delayed pouring on the accuracy of three different irreversible hydrocolloid impression materials - Regular set CA 37(Cavex, The Netherlands), regular set chromatic (Jeltrate, Dentsply), and fast set (Hydrogum soft, Zhermack Clinical) was investigated. A brass master die that contained two identical posts simulating two complete crown-tapered abutment preparations with reference grooves served as a standardized master model. A total of 120 impressions were made using specially prepared stock-perforated brass tray with 40 impressions of each material. The impressions were further sub-grouped according to four different storage time intervals: 0 min (immediately), 12 min, 30 min, and 1 h. The impressions were stored at room temperature in a zip-lock plastic bag. Interabutment and intraabutment distances were measured in the recovered stone dies (Type IV, Kalrock) using a profile projector with an accuracy of 0.001 mm. The data so obtained was analyzed statistically. Results: Results of this study showed no statistically significant differences in the accuracy of casts obtained at different time intervals. Conclusion: Because it is not always possible to pour the impression immediately in routine clinical practice, all irreversible hydrocolloid materials studied could be stored in a zip-lock plastic bag for upto 1 h without any significant distortion.

Keywords: Accuracy, delayed pouring, dimensional stability, irreversible hydrocolloids, profile projector


How to cite this article:
Wadhwa SS, Mehta R, Duggal N, Vasudeva K. The effect of pouring time on the dimensional accuracy of casts made from different irreversible hydrocolloid impression materials. Contemp Clin Dent 2013;4:313-8

How to cite this URL:
Wadhwa SS, Mehta R, Duggal N, Vasudeva K. The effect of pouring time on the dimensional accuracy of casts made from different irreversible hydrocolloid impression materials. Contemp Clin Dent [serial online] 2013 [cited 2019 Nov 18];4:313-8. Available from: http://www.contempclindent.org/text.asp?2013/4/3/313/118368


   Introduction Top


Impression making and pouring are critical steps in the process of producing successful dental prosthesis. Impression materials should reproduce hard and soft tissues accurately to obtain biologically, mechanically, functionally, and aesthetically acceptable restorations. [1] Accurate casts are required to perform multitude of functions like assisting in patient education, diagnosis of malocclusion, determination of the length and width of clinical crowns, identification of the need for surgical interventions, fabrication of custom trays, occlusal devices, and implant surgical guides. Hence, a precise, undistorted impression must be ensured for the accuracy of gypsum casts for correct treatment planning, constructing well-fitting removable and fixed prosthesis, and successful rehabilitation.

Alginate, an elastic, irreversible hydrocolloid impression material has been the staple of most dental practices for many years. The general use of irreversible hydrocolloid far exceeds that of any impression material, because of its various advantages such as hydrophilicity, pleasant taste and odor, non-staining, inexpensive, ease of mixing and effective use in the presence of saliva. Alginate is used to generate gypsum casts for numerous applications such as in making diagnostic casts, provisional crowns and bridges, orthodontic study models, sports mouth guards, bleaching trays, and fabrication of removable prosthesis. [2],[3],[4]

As with any hydrocolloid, alginates are approximately 85% water and are prone to distortion caused by expansion associated with imbibition (absorption of moisture), shrinkage due to moisture loss by evaporation, or continued reaction of the sol (syneresis). [5],[6],[7]] Due to host of contingencies, many dentists do not pour their own impressions immediately. Impression must be stable enough to produce accurate casts over extended periods of time. In general, the shorter this period, the higher is the accuracy. According to Morrow and colleagues, the most common error made in using alginate impression materials is not pouring the gypsum product into the impression immediately. [8] Various researchers and clinicians state that the casts produced from alginate impressions must be generated immediately or within 12 min after the impression is removed from the patient's mouth. Dahl et al., concluded that clinically acceptable working casts could be obtained from irreversible hydrocolloid impressions poured even after 3 h. [9] Recently, extended storage alginate impression materials have been marketed with claims that the materials exhibit dimensional stability same as that of elastomers and can be delayed poured even after 100 h. [10]

However, little information is provided in manufacturer's instruction sheet regarding the storage period and storage methods of alginate impression material. Thus, confusion continues to prevail regarding their storage time, and further research is required for a definite conclusion.

This study was thus undertaken to evaluate the effect of storage time on the dimensional accuracy of casts made from different irreversible hydrocolloids.


   Materials and Methods Top


Three different irreversible hydrocolloids used in the study were CA 37 (Cavex, The Netherlands), regular set; Jeltrate (Dentsply), regular set chromaticandHydrogum soft, Zhermack Clinical) - fast set. A brass master die of size 52 × 24 mm, containing two identical posts, simulating complete crown-tapered abutment preparations served as the standardized master model. The abutments were prepared with reference cross grooves on occlusal and proximal surfaces for reference measurements. Reference measurements [Figure 1] (made by profile projector with an accuracy of 0.001 mm or 1 μm) of the standardized model were as follows [Table 1]
Figure 1: Schematic representation of master model

Click here to view
Table 1: Dimensions of the brass master model

Click here to view


A specially prepared stock-perforated brass tray having a uniform space of approximately 4 mm for the irreversible hydrocolloid impression material was used to register the impression [Figure 2].
Figure 2: Brass master die and perforated stock tray

Click here to view


A rectangular brass plate was fabricated with a slit in the centre that allowed the handle of the master die to pass. This stabilized the master model on the index during impression making and thus standardized the positioning and thickness of the impression material [Figure 3].
Figure 3: Die stone slab, model index, perforated stock tray and master die

Click here to view


According to the manufacturer's recommended water/powder ratio for each material, a pre-weighed amount of alginate was mixed manually with distilled water at room temperature for indicated time periods. The brass master die that was maintained at 37°C to closely simulate the oral environment was removed from the oven and placed on the index just before making the impression. The alginate mix was loaded in the perforated brass tray and the impression was taken within the setting time as indicated by the manufacturer at room temperature [Figure 4]. The impressions were not rinsed with water or immersed in any disinfecting solution.
Figure 4: Tray placement during impression making of the master model

Click here to view


A total of 120 impressions were made using three different irreversible hydrocolloids with 40 impressions of each material. The 40 impressions of each material were further sub-grouped according to the four different storage time intervals to which these impressions were subjected, with 10 impressions into each time interval. The four different storage time intervals studied were 0 min (immediately), 12 min, 30 min, and 1h.

For the storage time indicated as 0 min, the casts were poured immediately after removal from the master model. Impressions, other than those that were poured immediately, were stored at room temperature in a zip-lock plastic bag in which a paper towel wetted with 50 ml of distilled water had been inserted 10 min earlier.

To control the amount of moisture, each paper towel was squeezed for 10 s between two glass slabs of size 13 × 13 cm 2 . The paper was positioned to avoid a direct contact with the tray and alginate [Figure 5].
Figure 5: Storage of the impression in a zip lock plastic bag

Click here to view


After the designated time interval, the impressions were poured with type IV die stone. The die stone was allowed to set for a minimum of 1 h before being separated from the impression. Casts obtained were numbered and subjected to measurements.

A profile projector with 15X magnification was utilized for the measurements of the stone casts. The following interabutment and intraabutment dimensions on the stone casts were measured and compared to the master model [Figure 6] and [Figure 7].
Figure 6: Die stone measurements with profile projector

Click here to view
Figure 7: Reference points and linear dimensions measured. AB -interabutment dimension, CD, EF, GH, IJ -interabutment dimension

Click here to view


Each stone cast measurement was repeated three times and the mean for all distance measurements was calculated. The readings thus obtained were put to statistical analysis.


   Results Top


The results of the study indicated that all the materials exhibited a continuous decrease in the interabutment (AB) distance with delay in pouring [Figure 8]. The mean difference in interabutment dimension ranged from 0.0143 mm at 0 min to 0.0531 mm at 1 h for the regular set Cavex, 0.0164 mm at 0 min to 0.0447 mm at 1 h for the regular set chromatic Jeltrate, and 0.0134 mm at 0 min to 0.0525 mm at 1 h for fast set Hydrogum.
Figure 8: Line diagram showing mean values for the interabutment distance (AB) for each group at four different time intervals

Click here to view


On the other side, an increase in each intrabutment dimensions measured (CD, EF, GH, and IJ) was observed with delay in pouring in all the three hydrocolloid impression materials [Figure 9],[Figure 10],[Figure 11],[Figure 12]. In all intra-abutment dimensions, the minimum increase from master value was observed in stone casts poured immediately, whereas maximum increase was observed in the casts poured at an interval of 1 h. The mean difference in intra-abutment dimensions ranged from −0.009 mm to −0.0746 mm for Cavex, −0.0168 mm to −0.0798 mm for Jeltrate, and −0.0075 mm to −0.0747 mm for Hydrogum.
Figure 9: Line diagram showing mean values for the intraabutment distance (CD) for each group at four different time intervals

Click here to view
Figure 10: Line diagram showing mean values for the intraabutment distance (EF) for each group at four different time intervals

Click here to view
Figure 11: Line diagram showing mean values for the intraabutment distance (GH) for each group at four different time intervals

Click here to view
Figure 12: Line diagram showing mean values for the intraabutment distance (IJ) for each group at four different time intervals

Click here to view


These observed changes showed that the immediate pouring produced the most accurate casts for all the irreversible hydrocolloid materials studied. The one-way analysis of variance (ANOVA) analysis showed that this decrease in interabutment dimension and the increase in intraabutment dimensions with time for each hydrocolloid material was statistically insignificant (P > 0.05).


   Discussion Top


Despite its numerous advantages, the dimensional accuracy of alginate-based impression material is one of the major concerns regarding the usage of this material. [1] The most common cause for error associated with the clinical and laboratory use of hydrocolloids is the distortion of the material following the removal of the impression from the mouth and the pouring of the cast. Hydrocolloid gel is approximately 80% water and is susceptible to dimensional changes when exposed to different environments. It may undergo expansion by absorbing water (imbibition) or shrinkage by losing water through evaporation and continued reaction of the sol (syneresis). [3],[6],[11] Imbibition, water evaporation, and syneresis may result in the production of inaccurate casts. The first two processes depend primarily on storage conditions, and syneresis is affected by the proprietary constituents of the alginate. [11],[12] In addition to water balance, alginate impression may also be distorted because of release of stresses which are induced by the application of pressure on the tray during the gelatinous stage. If the impression is not immediately poured, the stresses will become almost completely relaxed, resulting in serious distortions. [13],[14]

The present study was aimed at evaluating the effect of delayed pouring on the accuracy of three different irreversible hydrocolloid impression materials (regular set, regular set chromatic, and fast set). The objective was to determine if the hydrocolloid impression material would maintain their accuracy on being poured at the convenience of the operator over a reasonable time period.

Irreversible hydrocolloid exhibited a net contraction on conversion from sol to insoluble gel. The shrinkage of the material mostly occurred toward the centre of the mass or the bulk of the material. As the maximum bulk of the material was between the two posts, the shrinkage resulted in decrease of the interabutment distance, causing the centre of the two posts to come closer with delayed pouring. This measured change was similar to the change observed in the clinical situations. During full-arch impressions, the greatest amount of alginate was visibly located in the palatal zone, and the shrinkage occurred toward the larger bulk in the palatal region as compared to the other portions. [11],[15] Therefore, the palatal part of the casts would be smaller than the palatal part of the mouth. Although the major connector crossing the palate fit the cast well, they would not be snug against the palate in the mouth. [15],[16]

If the impression material was bonded firmly to the tray, shrinkage would result in the impression material being pulled toward the tray, causing an increase in die diameter. [17] As the material was firmly bonded to the perforated brass-stock tray, due to continued shrinkage, the impression material around the die was subjected to the centrifugal tensile forces, resulting in an increase in die diameter. This explained the increase in intraabutment distances that resulted in wider dies as compared to the master model.

All the impression materials studied had undergone shrinkage, possibly due to syneresis and evaporation of water. The amount of shrinkage caused by syneresis was always higher than evaporation. [18] In 100% relative humidity, the alginate materials first expanded and then shrink with time. [19] Maintenance of 100% relative humidity might lead to absorption of water from the humid environment (imbibition), resulting in distortion. [18],[20],[21] This could be avoided by storing the impressions in zip-sealed bags. To minimize the distortion caused by water changes in the irreversible hydrocolloid, the manufacturers add filler and smaller amounts of other proprietary ingredients to control the dimensional changes.

According to Thongthammachat et al., (2002) the deviation is clinically acceptable, if distortion occurred within the limit of the periodontal ligament space, which is in the range of 90-240 μm. [22],[23],[24],[25] In the present study, the deviation measured for all the impressions stored up to 1 h were well within this acceptable range, suggesting that the studied irreversible hydrocolloid impression materials could be used to produce clinically acceptable casts even on delayed pouring up to 1 h.

Rudd (2001) suggested that, by metal grinding, incorrectness of up to 150 μm in the cobalt chromium metal framework could be easily adjusted during try in stage for removable partial denture; therefore, this discrepancy might be considered acceptable. [26],[27],[28] This distortion is larger than the greatest dimensional change recorded for all the materials in the present study. This implied that even if the pouring of alginate is delayed by up to 1 h, it still produces stone casts that are clinically acceptable for partial denture framework construction.

Thus, the results of the present study suggested that all the three tested impression materials when stored properly for up to 1 h were dimensionally stable enough for fabrication of diagnostic casts, occlusal splints, acrylic appliances, and master casts for removable partial dentures. [29],[30] The decades old tenet that alginate impression materials must be poured immediately and wrapped in a damp towel or stored before pouring in gypsum may no longer be valid for every alginate impression, if it is stored adequately for limited times. As delayed pouring of alginate by up to 1 h did not significantly affect the dimensional accuracy of casts obtained, it might provide the clinician sufficient time to finish necessary chair-side procedures, leading to improved operator efficiency and clinical time management. Although the present in vitro study showed no statistically significant difference in the accuracy of casts obtained at different time intervals, it did not simulate the oral conditions such as effect of oral fluids, soft tissues, presence of undercuts, and different arch forms. The present study evaluated only the effect of delayed pour on the dimensional accuracy of alginate impression material. There are various other factors that may influence the precision of dental cast such as type of impression material, various impression techniques, type of trays used, the materials used for making casts, and various disinfectant procedures to which impressions are subjected. All these factors need further investigation under more closely simulated clinical conditions.

 
   References Top

1.Faria AC, Rodrigues RC, Macedo AP, Mattos Mda G, Ribeiro RF. Accuracy of stone casts obtained by different impression materials. Braz Oral Res 2008;22:293-8.  Back to cited text no. 1
[PUBMED]    
2.Carr AB, McGivney GP, Brown DT. Impression materials and procedures for removable partial dentures. McCracken's Removable partial prosthodontics, 11 th ed. St. Louis: Mosby; 2008. p. 273-4.  Back to cited text no. 2
    
3.Nandini VV, Venkatesh KV, Nair KC. Alginate impressions: A practical perspective. J Conserv Dent 2008;11:37-41.  Back to cited text no. 3
[PUBMED]  Medknow Journal  
4.Sedda M, Casarotto A, Rausita A, Borracchini A. Effect of storage time on the accuracy of casts made from different irreversible hydrocolloids. J Contemp Dent Pract 2008;9:59-66.  Back to cited text no. 4
    
5.Jamani KD. The effect of pouring time and storage condition on the accuracy of irreversible hydrocolloid impressions. Saudi Dent Journal 2002;14:126-30.  Back to cited text no. 5
    
6.Rubel SB. Impression materials: A comparative review of impression materials most commonly used in restorative dentistry. Dent Clinc North Am 2007;51:629-42.  Back to cited text no. 6
    
7.Walker MP, Burckhard J, Mittis DA, Williams KB. Dimensional change over time of extended storage alginate impression materials. Angle Orthod 2010;80:1110-5.  Back to cited text no. 7
    
8.Morrow RM, Rudd KD, Eissmann HF. Preliminary impressions: Care and pouring. In: Dental Laboratory Procedures: Complete dentures, Vol. 1. St. Louis: Mosby; 1980. p. 1-17.  Back to cited text no. 8
    
9.Dahl LB, Dymbe B, Valderhaug J. Bonding properties and dimensional stability of hydrocolloid impression systems in fixed prosthodontics. J Prosthet Dent 1985;53:796-800.  Back to cited text no. 9
    
10.Jaminsika J, Vandewalle K. Dimensional stability of alginate impression materials. IADR General Session; 2009.  Back to cited text no. 10
    
11.Imbery AT, Nehring J, Janus C, Moon CP. Accuracy and dimensional stability of extended-pour and conventional alginate impression materials. J Am Dent Assoc 2010;14:32-9.  Back to cited text no. 11
    
12.Buchan S, Peggie RW. Role of ingredients in alginate impression compounds. J Dent Res 1966;45:1120-9.  Back to cited text no. 12
[PUBMED]    
13.Rudd KD, Dunn WB. Accurate removable partial dentures. J Prosthet Dent 1967;18:559-70.  Back to cited text no. 13
    
14.Rudd KD, Morrow MR, Bange AA. Accurate casts. J Prosthet Dent 1969;21:545-54.  Back to cited text no. 14
    
15.Rudd KD, Morrow MR, Strunk RR. Accurate alginate impressions. J Prosthet Dent 1969;22:294-300.  Back to cited text no. 15
    
16.Kaiser DA, Nicholls JI. Astudy of distortion and surface hardness of improved artificial stone casts. J Prosthet Dent 1976;36:373-81.  Back to cited text no. 16
[PUBMED]    
17.Skinner EW, Hobilt NE. A study of the accuracy of hydrocolloid impressions. J Prosthet Dent 1956;6:80-6.  Back to cited text no. 17
    
18.Reddy CG, Gujjari KA, Shambhu HS. Influence of long term storage on the linear dimensional accuracy of improved alginate impressions-an in vitro study. TPDI 2010;1:34-7.  Back to cited text no. 18
    
19.Hiraguchi H, Kaketani M, Hirose H, Yoneyama T. The influence of storing alginate impressions sprayed with disinfectant on dimensional accuracy and deformation of maxillary edentulous stone models. Dent Mater J 2010;29:309-15.  Back to cited text no. 19
[PUBMED]    
20.Alcan T, Ceylanoglu C, Baysal B. The relationship between digital model accuracy and time dependent deformation of alginate impressions. Angle Orthod 2009;79:30-6.  Back to cited text no. 20
    
21.Coolidge ED. The thickness of the human periodontal membrane. J Am Dent Assoc 1937;24:1260-70.  Back to cited text no. 21
    
22.Thongthammachat S, Moore BK, Barco MT 2 nd , Hovijitra S, Brown DT, Andres CJ. Dimensional accuracy of dental casts: Influence of tray material, impression material and time. J Prosthodont 2002;11:98-108.  Back to cited text no. 22
    
23.Farzin M, Panahandeh H. Effect of pouring time and storage temperature on dimensional stability of casts made from irreversible hydrocolloid. J Dent (Tehran) 2010;7:179-84.  Back to cited text no. 23
[PUBMED]    
24.Rudd RW, Rudd KD. A review of 243 errors possible during the fabrication of a removable partial denture: Part II. J Prosthet Dent 2001;86:262-76.  Back to cited text no. 24
[PUBMED]    
25.Peutzfeldt A, Asmussen E. Accuracy of alginate and elastomeric impression materials. Scand J Dent Res 1989;97:375-9.  Back to cited text no. 25
[PUBMED]    
26.Miller WM. Syneresis in alginate impression materials. Br Dent J 1975;139:425-30.  Back to cited text no. 26
    
27.Harris WT Jr. Water temperature and accuracy of alginate impressions. J Prosthet Dent 1969;21:613-7.  Back to cited text no. 27
[PUBMED]    
28.Fusayama T, Hosoda H. Distortion of irreversible hydrocolloid and mercaptan rubber-base impressions. J Prosthetic Dent 1961;11:318-33.  Back to cited text no. 28
    
29.Fairchild JM. Versatile uses for alginate impression material. J Prosthet Dent 1974;31:266-9.  Back to cited text no. 29
[PUBMED]    
30.Supowitz ML, Schnell RJ, Dykema RW, Goodacre CJ. Dimensional accuracy of combined reversible and irreversible hydrocolloid impression materials. J Prosthet Dent 1988;59:404-9.  Back to cited text no. 30
[PUBMED]    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7], [Figure 8], [Figure 9], [Figure 10], [Figure 11], [Figure 12]
 
 
    Tables

  [Table 1]


This article has been cited by
1 Three dimensional accuracy analysis of dental stone casts fabricated using irreversible hydrocolloid impressions
Young-Hun Joo,Jin-Han Lee
Journal of Dental Rehabilitation and Applied Science. 2015; 31(4): 316
[Pubmed] | [DOI]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Add to My List *